Description
The vial contains a mixture of 3.0 to 3.4 µm rainbow particles that are stained with eight different fluorescent intensities. Every particle of the rainbow contains a mixture of fluorophores that are excited at any wavelength from 365 to 650 nm. Rainbow particles have emission spectra compatible with many common fluorophores used for immunofluorescent staining with flow cytometric analysis.
Preparation and storage
Store undiluted at 4 ° C and protected yourself from prolonged exposure to light. Do not freeze.
Application notes
Recommended test procedure:
- This particle mixture (~ 10×10 ^ 6 particles / ml) is useful for routine calibration of flow cytometers. Before use, resuspend the particles by vortexing.
- Dilution of 3 to 5 drops of particles in 1 ml of sheath fluid will provide an adequate number of particles for flow cytometric analysis.
Product notices
1. Since applications vary, each investigator should titrate the reagent for optimal results.
Caution: Sodium azide produces highly toxic hydrazoic acid under acidic conditions. Dilute azide compounds in tap water before disposal to prevent accumulation of potentially explosive deposits in plumbing.
2. Cy is a trademark of GE Healthcare.
3. See www.bdbiosciences.com/pharmingen/protocols for technical protocols.

Rainbow calibration particles (6 peaks)
Each vial contains a mixture of rainbow particles that are stained with six different fluorescent intensities.
Rainbow calibration particles (8 peaks)
The vial contains a mixture of rainbow particles that are stained with eight different fluorescent intensities.
Rainbow fluorescent particles (high intensity)
Each vial contains a single population of rainbow particles that stain with a single fluorescent intensity. The intensity of this particle corresponds to the brightest intensity of the peak in the rainbow calibration particles, 3.0-3.4μm (Ref. No. 556286)
Rainbow Fluorescent Particles (FL1 Midrange Fluorescence)
Each vial contains a single population of rainbow particles that stain with a single fluorescent intensity.
Blank calibration particles
- Each vial contains a single population of blank calibration particles that are stained with a single fluorescent intensity.
- The intensity of this particle corresponds to the fainter intensity peak in the rainbow calibration particles (6 peaks), 6.0-6.4μm (Ref. No. 5556288)
- Approximately the background fluorescence of unstained cells